Initial Setup |
Observing efficiency |
| Imaging |
10 mins |
25% (13% in Qa filter) |
| Imaging polarimetry |
10 mins |
15% (11% in Qa filter) |
| Chop/nod mode spectroscopy |
15 mins |
25% |
| Stare mode spectroscopy |
15 mins |
67% |
Mid-IR observing overheads can be significant and
must be taken into account when proposing for time on Michelle.
Experience indicates that each new imaging target will incur a
configuration overhead of 10 minutes (e.g. slewing, centering, setting
the peripheral wavefront sensor for tip-tilt AO correction on a guide
star). For setting up a spectroscopic measurement an additional 5
minutes should be allowed for accurate centering on the slit and for
switching between imaging and spectroscopic modes.
Once a typical observation with chopping and nodding (i.e, imaging
and low-resolution spectroscopy) begins, the time spent actually
integrating on the source - the value entered in the observing tool as
the "total on-source time" - will be about 25% of the elapsed
time. Thus, you specify the actual on-source time for Nod A Chop A +
Nod B Chop B (i.e., one beam for each nod position). If chopping
on-chip you in fact get double the on-source time that you specified
but, because guiding is only possible in one of the chop beams (see here), the quality of one of the
two images will be somewhat degraded and unlikely to be useful for
science. It is not necessary (or, indeed, possible) to specify the
time for a single exposure of the array, or the time (and number of
individual exposures) between nods - this is done internally and
depends on the instrumental configuration. You only need be concerned
with the total on-source time and total elapsed time. The factor of
two on-source overheads are due to the chop duty cycle and the time
required for the telescope to nod and then stabilise. We hope that
this eventually can be reduced to a factor of ~1.5.
The on-source efficiency for imaging with the Qa filter is about
half that of any of the other imaging filters. This is due to
increased internal overheads that are required in Q-band where the
background is much higher than it is in N-band. It is necessary to
chop faster for the Qa filter than for the N-band filters, and the
smaller time per chop position combined with larger detector overheads
for reading out in this configuration means that the number of frames
that can be taken per chop position turns out to be much smaller than
it is for any of the other filters - hence the significant reduction
in the on-source efficiency.
When observing with the echelle and medium resolution
gratings, which are performed in stare/nod (as opposed to
chop/nod) mode, the observing overhead, after the initial 15 minutes
of setup time, is a factor of about 1.5.
In the case of imaging
polarimetry the time requested in the observing tool (and the
time that should be specified in the integration time calculator) is
the time per wave-plate position, so that the total integration
time on target is four times that which is requested. The
overhead involved in moving the half-wave plate during the
observations means that the overall efficiency is generally of order
15%, rather than 25% as in regular imaging mode. Thus for a 1 minute
on-source time specified in the OT, the actual time spent on-source
will be 4 minutes; coupled with the 15% efficiency, the total time
needed to take the observation will be about 4 minutes/0.15 = 27
minutes for the filters in the N-band window (after acquisition). For
the Qa filter the time required is a factor of 1.4 times longer than
it is for the N-band filters. The on-source efficiency in that case
is about 11% (which turns out to not be that much smaller than in
regular imaging mode).
If one prefers to think of the efficiency in terms of the total
time on-source specified in the OT, then the efficiency should be
taken as 15% / 4 waveplate positions = 3.75% for the filters in the
N-band window. For Qa the value is 11% / 4 waveplate positions =
2.8%.
|