The following tables provide proper GCAL configurations and exposure times for GCAL flats and arcs for GNIRS.
The values given cover a range of configurations, slit widths and
wavelengths.
If your particular configuration is not here, you may be able to extrapolate
from similar settings. If not,
please submit a query via the HelpDesk.
UPDATE May 2006: New GNIRS CAL sensitivities are now listed below
following the installation of a new NIR color balance GCAL filter.
Sufficient illumination of XD orders 3 - 9 (2.55 - 0.77 microns)
is now achieved with 2 GCAL flat settings without saturating the lower orders.
If your science needs to utilize XD order 10, then please specify this
in a NOTE within your phase II, and let your contact scientist know. They
will add a third flat field field step to the template sequence below.
UPDATE March 2005: New xenon (Xe) lamp REPLACES krypton (Kr) lamp to provide better
wavelength calibration for R=18,000 observations.
Long slit, short camera, 111 l/mm grating (LS-S111)
Long slit, long camera, 111 l/mm grating (LS-L111)
Cross-Dispersed 32 l/mm, Short Camera (Long Camera not used with XD-32)
GNIRS Configuration
GCAL Configuration
Filter
Slit Width (arcsec)
Lamp
Filter
Exposure Time (s)
Coadds
XD
0.3, 0.45, 0.675
IR*
NIR balance
1.5
8
XD
1.0, 3.0
IR*
NIR balance
0.2
8
XD
0.3, 0.45, 0.675
QH**
NIR balance
0.6
9
XD
1.0, 3.0
QH**
NIR balance
0.2
9
XD
0.3, 0.45, 0.675, 1.0, 3.0
Ar
none
0.2
20
Cross-Dispersed 111 l/mm, Short Camera (Long Camera not yet available with XD-111)
GNIRS Configuration
GCAL Configuration
Filter
Slit Width (arcsec)
Lamp
Filter
Exposure Time (s)
Coadds
XD
0.3, 0.45, 0.675
IR*
NIR balance
4.0
4
XD
1.0, 3.0
IR*
NIR balance
0.6
8
XD
0.3, 0.45, 0.675
QH**
NIR balance
1.5
6
XD
1.0, 3.0
QH**
NIR balance
0.2
8
XD
0.3, 0.45, 0.675, 1.0, 3.0
Ar
none
0.2
20
* The IR lamp is a better flat field source for XD Order 3 (the first continuous order viewed from the left in the image display).
** The QH lamp now illuminates ALL the XD orders without saturating the array! There is an absorption feature due to the QH source present in the first order on the left as seen in the display (XD order 3). Use the IR lamp flat for this order only instead of the QH flat.
IFU 32 l/mm, Short Camera (Long Camera not used with IFU-32)
GNIRS Configuration
GCAL Configuration
Filter
Lamp
Filter
Exposure Time (s)
Coadds
K
IR
NIR balance
4.0
2
H
QH
NIR balance
3.0
2
J
QH
NIR balance
10.0
3
K
Ar
none
10
2
H
Ar
none
2.5
2
J
Ar
none
1.2
2
IFU 111 l/mm, Short Camera (Long Camera not used for IFU 111)
GNIRS Configuration
GCAL Configuration
Filter
Lamp
Filter
Exposure Time (s)
Coadds
K
IR
NIR balance
10.0
3
H
QH
NIR balance
7.0
2
J
QH
NIR balance
10.0
4
K
Ar
none
20
2
H
Ar
none
10
2
J
Ar
none
5.0
4
Long Slit 32 l/mm, Short Camera
GNIRS Configuration
GCAL Configuration
Wavelength
Slit Width (arcsec)
Lamp
Filter
Exposure Time (s)
Coadds
J
0.3, 0.45, 0.675
QH
NIR balance
2.0
3
J
1.0, 3.0
QH
NIR balance
0.6
3
H
0.3, 0.45, 0.675
QH
NIR balance
0.6
4
H
1.0, 3.0
QH
NIR balance
0.2
3
K
0.3, 0.45, 0.675
IR
NIR balance
1.0
5
K
1.0, 3.0
IR
NIR balance
0.2
5
L, M
0.3, 0.45, 0.675
blocked
NIR balance
0.2
12
J
0.3, 0.45, 0.675, 1.0, 3.0
Ar
none
0.6
2
H
0.3, 0.45, 0.675, 1.0, 3.0
Ar
none
0.6
3
K
0.3, 0.45, 0.675, 1.0, 3.0
Ar
none
4.0
3
Long Slit 111 l/mm, Short Camera
GNIRS Configuration
GCAL Configuration
Wavelength
Slit Width (arcsec)
Lamp
Filter
Exposure Time (s)
Coadds
J
0.3, 0.45, 0.675
QH
NIR balance
5.0
3
J
1.0, 3.0
QH
NIR balance
1.0
3
H
0.3, 0.45, 0.675
QH
NIR balance
1.0
6
H
1.0, 3.0
QH
NIR balance
0.2
6
K
0.3, 0.45, 0.675
IR
NIR balance
2.5
6
K
1.0, 3.0
IR
NIR balance
0.5
6
L, M
0.3, 0.45, 0.675
blocked
NIR balance
0.4
12
J
0.3, 0.45, 0.675, 1.0, 3.0
Ar
none
0.4
5
H
0.3, 0.45, 0.675, 1.0, 3.0
Ar
none
0.6
5
K
0.3, 0.45, 0.675, 1.0, 3.0
Ar
none
10.0
3
Long Slit 111 l/mm, Long Camera
GNIRS Configuration
GCAL Configuration
Wavelength
Slit Width (arcsec)
Lamp
Filter
Exposure Time (s)
Coadds
J
0.1, 0.2, 0.3
QH
NIR balance
1.0
8
H
0.1, 0.2, 0.3
QH
NIR balance
1.0
8
K
0.1, 0.2, 0.3
IR
NIR balance
3.0
6
L
0.1, 0.2, 0.3
blocked
none
10.0
6
M
0.1, 0.2, 0.3
blocked
none
4.0
5
J
0.1, 0.2, 0.3
Ar, Xe*
none
20
4
H
0.1, 0.2, 0.3
Ar, Xe*
none
20
4
K
0.1, 0.2, 0.3
Ar, Xe*
none
20
4
*Both Ar and Xe lamp exposures may be needed for sufficient
calibration in some wavelength regions at this resolution. Check for
availability of Ar and Xe emission lines here.