NIRI Optical Overview
|
|
The NIRI beam-splitter wheel contains three beam-splitters to direct
the science beam and OIWFS (On-instrument wavefront sensor) beam to
the appropriate on-board instruments. The beam-splitters are flat
mirrors of different sizes to allow the OIWFS beam for each camera,
f/6, f/14 and f/32, to be picked off without modifying the science
field. The beam-splitters therefore, send the available field
(dependent on the science camera in use) surrounding the science
field, to the OIWFS which operates in the near-IR using a Rockwell
1024x1024 near-IR array detector. The photo below shows the
beam-splitter wheel in-situ in the NIRI dewar. The wheel holds three
beam-splitters (one to the right, one to the left and one at the
bottom of the wheel) and an alignment pin-hole mask for the OIWFS
beam.
The photo below shows the entrance window at the bottom and the f/32 beam-splitter mirror located over the focal-plane pin-hole mask in the focal-plane mask wheel. To the left of the f/32 mask is a larger beam-splitter, the f/6 unit, and to the right is the third beam-splitter, the f/14 unit. All three beams-splitters are mounted on the beam-splitter wheel.
Below is shown a closer view of the f/32 beam-splitter. The telescope beam enters the photo from the bottom where it is deflected 90 degrees by the beam-splitter (into the photo). Next, the beam passes through the focal-plane mask wheel set, in this photo, to the pin-hole mask position.
|
Last update July 31, 2000; Colin Aspin