[NIRI internal CAD]

NIRI Grisms

NIRI contains several KRS direct-ruled grisms (prisms with transmission gratings on one surface) and slits of various widths and offsets from the center of the field of view. Their combinations permit low to moderate resolution spectroscopy through most of the 1-5um region. The first table below gives predicted spectral resolving powers and wavelength coverages. Only one wavelength range and slit width (or spectral resolution) is configurable at one time

The available slit widths at f/6 are 0.232 and 0.464 arcsec (2 and 4 pixels) wide; at f/14 the slits are 0.10 and 0.15 arcsec (2 and 3 pixels) wide. See the list of available slits for details and estimates of throughput as a function of image quality. The listed resolving powers are at the central wavelengths of the accessible wavelength ranges. It is assumed that for all available slit widths the target overfills the slit.

Initial tests indicate that in the J, H, and K bands resolving powers do not match those expected for ideal grisms. Little information is available concerning the performance of the L and M grisms, except that they work at shorter wavelengths in second or third orders. The performances of these grisms may also be degraded from the ideal and we have assumed so in the tables below. The tables will be updated as soon as new information becomes available.

At f/6 The accessible wavelength ranges are limited either by the array size, the blocking filters (whose properties are given in the second table below) or the atmospheric window. In general, with the f/6 camera full atmospheric windows are spectrally imaged onto the detector.

At f/14 the wavelength ranges are 2.3 times narrower. The resolutions, where measured, are comparable to those measured at f/6. No significant increase in resolution was found with the J, H, and K grisms at f/14. This means that at f/14 a resolution element falls across roughly 2.3 times as many pixels as at f/6. Sensitivities will be reduced accordingly. Except perhaps at L and M the only valid scientific justification for using the grisms at f/14 is to obtain higher angular resolution. As mentioned, the detector sees a wavelength range that is considerably less than the atmospheric window. With the centered 2-pixel and 3-pixel wide slits at f/14 the center of the atmospheric window is centered on the detector. The other 2-pixel and 3-pixel wide slits, offset to either side of the center, cover wavelength ranges offset to the blue and red. The wavelength ranges are approximate; accurate values will be posted as soon as they are measured

Waveband Camera Wavelength Range
(microns)
Slit Width
(arcsec)
Resolving Power
at center of range
(l/dl)
Order-sorting filter transmission curve available
J f/6 1.02 - 1.40 0.232 or 0.464 850 or 600 [check/tick mark]
0.97 - 1.35
f/14 1.13 - 1.29 0.100 or 0.150 800 or 700
1.04 - 1.19
1.18 - 1.34
H f/6 1.48 - 1.78 0.232 or 0.464 1200 or 750 [check/tick mark]
f/14 1.53 - 1.76 0.100 or 0.150 1300 or 1200
1.48 - 1.66
1.64 - 1.78
K f/6 1.90 - 2.49 0.232 or 0.464 950 or 700 [check/tick mark]
f/14 1.90 - 2.21 0.100 or 0.150 1100 or 1000
2.04 - 2.35
2.19 - 2.49
L f/6 3.01 - 4.13 0.232 or 0.464 1100 or 700 -
2.90 - 4.05
f/14 3.10 - 3.60 0.100 or 0.150 1500 or 1200
3.33 - 3.83
3.56 - 4.06
M f/6 4.45 - 5.65 0.232 or 0.464 1200 or 800 [check/tick mark]
f/14 4.45 - 5.02 0.100 or 0.150 1600 or 1300
4.45 - 5.35
4.97 - 5.65



The following table gives wavelength ranges of the order-sorting filters in NIRI that are used with the grisms and also lists the physical properties of each grism.

Waveband Central Wavelength (microns) Order-sorting Filter bandpass (microns) Grism Properties
lines/mm Prism angle
J 1.30 0.97-1.63 312.6 15
H 1.65 1.48-1.78 223.9 15
K 2.20 1.90-2.49 166.7 15
L 3.50 2.90-4.25 104.1 15
M 5.00 4.40-6.00 72.8 15


[Science Operations home] [NIRI home]



Last update August 24, 2000; Tom Geballe