NIFS Baseline Calibrations 

For all queue observations, a set of standard calibrations (the "baseline calibrations") will be taken by Gemini Staff to ensure the long-term utility of data in the archive. The baseline calibration set varies from instrument to instrument and from mode to mode.

Applicants should define baseline calibrations in their phase II files (by copying the appropriate one from the OT Library). If no baseline calibration was defined, then a delivery of such data won't be guaranteed. If additional calibrations are deemed necessary by applicants, for example to achieve a precision beyond that achievable via the baseline set, then these must be included explicitly as part of the Phase I proposal and Phase II science program. These additional calibration observations will be charged to the PI.

Any time used between the hours of nautical sunset and sunrise to obtain baseline calibrations will not be explicitly charged to the program but will be charged to the partner country whose NTAC recommended time (i.e. it is part of the time accounting used to balance partner usage). The time will be charged in proportion to the number of programs that receive the calibration data. 

Calibration Observations of NIFS

Measurement Type Notes
Bad pixel mask Can be derived with each dataset from GCAL flats and short darks. 
Calibration unit flat field GCAL flats* are taken nightly before or after each program in each wavelength setting. Closed shutter flats are subtracted to remove dark current and thermal emission from the camera optics.
Dark Highly suggested to take. NIFS raw images have many hot pixels which can be removed effectively by subtracting darks. Therefore, PIs are advised to define dark frames at the same exposure time as the science observations and include them in their phaseII files. These darks will only be acquired at the end of an observing night.
Wavelength calibration Argon (and/or Xenon) arc lamp measurements are made for each spectral setting. One arc and one corresponding dark frame will be obtained at the time of the science observations. Because of non-repeatability of NIFS grating positions, taking one arc (and a dark) before next grating change is required. Due to the nature of queue observation, PI will not know when the next grating change will happen. Therefore, it is recommend to define an arc and a dark for all NIFS observation. For a proper setup, see the OT Library.
Telluric standard star An appropriate telluric standard should be chosen by the PI and included in the Phase II program before and after each observation. When the observation is executed in the queue, the astronomer will select one of the two for the Baseline calibration. The PI will not be charged for one telluric standard for each observation. The PI should take care to select a star of the appropriate spectral type and brightness at an airmass that will match the science target. For help selecting telluric standards, see the spectroscopic standards page.
Flux standard star Not included as part of the baseline calibration set. Please include spectrophotometric standards in your proposal if needed.
Atmospheric extinction Not included as part of the baseline calibration set.
Point Spread Function star Not included as part of the baseline calibration set.
Calibration mask image for spatial rectification Part of the baseline calibration. Each PI science program should define one "Ronchi" Calibration mask image for each wavelength setting. The Ronchi calibration spectra are used to rectify the spatial scale and remove S-distortion from NIFS spectra.
Special standards Radial velocity or Line Index standards are not included as part of the baseline calibration set.

 

* IMPORTANT NOTE regarding GCAL flats and flatfield accuracy:

Flats taken with the calibration unit make use of shutter open and closed images to explicitly correct for dark current and thermal emission from the camera optics. The calibration unit produces a beam that matches the telescope pupil very nicely except that it contains no central obscuration. The light path between GCAL and NIFS obviously excludes the primary and secondary mirrors as well. The GCAL illumination is therefore subtly different from that of NIFS on the sky.

GCAL flats made with the NIRI infrared camera are reproducible from night to night to about 0.3%, ie, the sensitivity of a given pixel varies by 0.3% over many nights as measured by GCAL. We assume a similar stability will hold true for NIFS. Obviously, for a star that subtends many pixels, the photometric accuracy will be approximately 0.3% divided by the square root of the number of pixels.