T-ReCS Sensitivity

caution Important caveat: These values are based on calculations and on-telescope experience with OSCIR. They have not been confirmed experimentally with T-ReCS itself.

All sensitivities are 5 sigma in 1 hr. Overheads are not included i.e. a combined efficiency for chopping and nodding of 50% implies an elapsed time of 2 hr to obtain 1 hr of data.

Mode Point Source
(mJy)
Extended Source
(mJy / pixel)
Notes
Imaging
broad-band N (10.8µm) 1.0 0.14 (a, b)
narrow-band 9.8µm 2.2 0.31 (a, c)
broad-band Q (18µm) 21.5 1.8 (d)
Spectroscopy
low res. (10.8µm) 11 - (e)
medium res. (10.8µm) 32 - (e)
  1. Assumes combined system emissivity of telescope, warm optics and atmosphere of 8%, expected median delivered image quality (50% EED) for point source of about 0.4" FWHM and 0.7" diameter software aperture.
  2. Broad N-band filter FWHM = 5.3 µm.
  3. Narrow-band filter FWHM = 1.0 µm.
  4. Qa filter FWHM = 1.5 µm
  5. Spectroscopic sensitivities estimated by scaling from imaging benchmark as sqrt(relative resolving power / relative throughput); assumes 0.34 arcsec slit.