In this section we present recommended observing strategies for
several types of T-ReCS observing programs. The goal of defining these
strategies is to illustrate what observing procedures and calibrations
are needed to reach desired scientific goals.
We therefore recommend that all programs follow one of the strategies
listed below. If an observing plan does not match one of the
strategies, then Gemini staff should be contacted for advice.
Important note: These strategies are based on estimates
of the T-ReCS performance. All of the issues will be evaluated
during the T-ReCS commissioning.
Strategic issues:
The strategies were developed based on several issues.
Field of View: The T-ReCS field of view is about 20x25".
Chopping: With T-ReCS the typical chopping amplitude will be 20"
and the maximum about 25". The chop angle can be varied between 0 and
180 degrees east of north); the default will be 0 (North-South).
Observations in crowded fields require careful consideration of the
chop angle in order to avoid chopping onto other sources that may
surround the desired science target.
Nodding: The nodding mode for the first T-ReCS semester will
be beam-switched, in which the nodding amplitude and direction
are equal to the chopper settings.
Integration Time & Overheads:
Photometric Calibration: Gemini will take a baseline calibration
set consisting of a few measurements of bright standard stars spaced
through the night in order to determine approximate (10%) photometric
calibration and image quality, but NOT airmass corrections. See XXX
for details.
PSF Calibration: The baseline calibration data may also
be useful for an approximate characterization of the point-spread-function.
However, the Gemini PSF depends on many factors such as wind speed
and guide star brightness that may vary significantly during a night.
Baseline calibration data are not likely to be suitable for detailed
PSF comparisons or deconvolutions.
Flat Fielding: Mid-infrared flat-fields are typically derived
from data take on-sky at a few different airmasses, so they are
relatively expensive in terms of telescope time. We expect that the
T-ReCS detector pixels will be uniform in responsivity to 1-2% rms, and
that many T-ReCS observations will not be of sufficient signal-to-noise
(> 50-100 per pixel) to reveal these flat field effects. In these
cases, highly accurate flats will not improve the quality of the
science data; the data may be used with no flat fielding, or by
applying a recently-constructed flat obtained from a library.
If the source is very bright and extended (e.g. Mars, or the
Kleinmann-Low nebula in Orion), and high S/N flats are required,
then the requirements should be stated in the observing proposal and
suitable time should be requested to take the flats.
Astrometry: We expect the default pointing accuracy as indicated
in the WCS to be 1 to 2 arcsec. If better accuracy is desired,
Strategies
Simple Imaging
Example : Initial "reconnaisance" imaging of small targets.
Photometry: 10%
PSF : Low
Flat Field: Low
Astrometry: Low
The Simple Imaging strategy is suitable for quick snapshots of a large
number of targets, or of a few targets in many filters, when the
science goal is to determine approximate source positions, fluxes,
and morphologies. In this case the photometric and PSF calibrations
can be derived from the Baseline calibrations, the flat fields if needed
can be taken from a library, and astrometry can be derived from the
approximate (1") WCS in the FITS header and by registering the images
with radio or NIR data.
Mosaics
Example : Initial "reconnaisance" imaging of large targets.
Photometry: 10%
PSF : Low
Flat Field: Low
Astrometry: Low
This is similar to Simple Imaging, except that the field is larger
than the T-ReCS detector. In this case the position of each sub-field
must be specified, with due consideration of the desired overlap.
The chopping limitation of about 25" must also be considered; it
is generally difficult to make mosaics of large objects because one
cannot chop entirely off-source.
Deep Imaging of faint targets
Example: A two-hour integration on a high-redshift galaxy
in order to measure the N-band flux.
Photometry: 5%
PSF : Low
Flat Field: Low
Astrometry: High
Faint targets require long integration times with minimal
interruptions. If the final S/N is < 50 accurate flat fields
are not a priority. If better than baseline calibration is required,
special calibrations should be requested. In addition, the astrometry
can be well determined by observing a pointing offset star. In this
case, a suitable strategy would be:
Calibrate on standard star (specify which star)
Slew to pointing offset star near galaxy (bright enough to see on
T-ReCS). Place star on AcqCam hotspot, absorb offsets, take 1 T-ReCS
image to document position on array.
Slew to Galaxy, integrate 2 hours.
Calibrate on standard star.
High angular resolution imaging
Example : Imaging of T-Tauri star to detect extended disk emission.
Photometry: 10%
PSF : High
Flat Field: High
Astrometry: Low
In this case the photometric accuracy and astrometry requirements are
low because the star's total flux and position are already well known.
The PSF characterization is critical because the goal is to separate
faint disk emission from the bright, point-like stellar photosphere.
The flat field requirement is also important: even though the S/N
of the disk is low, the star's S/N is high enough so that accurate
flats may help separate its signal from the disk's.
Take flat fields.
Slew to PSF reference star, guide with P1 if possible for highest
stability. Integrate to specified S/N. A
typical integration time would be 5-10 minutes.
Slew to science target, integrate to specified S/N (5-10 min)
Repeat previous two steps for desired number of cycles.
Dither by 2" between every cycle.
Take another flat field set at end.
The optimal integration times and number of cycles are dependent on
the system stability and target brightness, and will be evaluated as
part of the T-ReCS commissioning and system verification.
Tom Hayward
Last modified: Sun Feb 17 17:46:48 CLST 2002