You are in: Instruments > Altair > Performance and Use > Performance Estimates   

[altair logo]

NIRI + Altair Performance Estimates 


Using the Altair + NIRI Integration Time Calculator we have estimated the Strehl ratio and point source brightness that gives a signal-to-noise ratio (S/N) or approximately 5 in a one-hour integration. See the commissioning results for examples of actual delivered performance.

The details of how the ITC calculates limiting magnitudes is described in the "more info" links accessible from the ITC. ITC help pages include information on the filter zero points, Altair properties, NIRI camera and detector properties, and calculation methods.

The ITC includes adjustments for observing conditions in its calculations. The meanings of the observing condition criteria are explained in detail in the Observing Conditions ITC help pages. For the tables presented here, we assumed photometric conditions (50%-ile cloud cover), the normal low sky background (80%-ile) and an airmass of less than 1.2. 

All magnitudes listed below result in a S/N of 5 in a total integration time of 1 hour. An individual exposure time of 600s was assumed for all cases. Observations are assumed to be dithered and sky subtraction noise is included. All individual exposures include the source. The ITC calculates the S/N in an aperture that maximizes S/N given the predicted image quality for the observing conditions and wavelength requested. The sensitivity values in the table below use the optimum ITC aperture. 

The selected conditions (image quality, guide star brightness and separation) are only intended to be illustrative examples.

Filter Image Quality 
%-ile Bin
Guide Star
R-band (mag)
Guide Star
Distance (arcsec)
Point Source 
Sensitivity (mag)
Strehl Ratio
J 20%-ile 12 5 24.0 0.115
70%-ile 12 5 22.5 0.015
15 5 21.9 0.001
H 20%-ile 12 5 23.9 0.318
70%-ile 12 5 22.8 0.102
15 5 21.7 0.019
K' 20%-ile 12 5  23.8 0.533
70%-ile 12 5  23.2 0.275
 15  22.4 0.126
15 5  22.2 0.100
 15  21.7 0.046
85%-ile 12 5  22.1 0.091
H2 1-0 S(1) 70%-ile 12 5  21.8 0.276

[Science Operations home] [Altair home]


Last update July 2, 2003; Francois Rigaut and Phil Puxley